PSØ21 resolution at 3 V Vio (Strain gage & PSØ21io) and 3 V (PSØ21core)

(with SINC3 Filter, for Fast-settle multply noise by 1.73)

Measuring rate in Hz
ENOB
No. of LSB @ 2 mV/V
Noise
SG Ratio2 mV/V3.3 mV/VµVppm
10,000
20.5
11.5
12.2
2,850
2.100
0.700
4,000
22.0
13.0
13.7
8,000
.750
0.250
2,000
22.5
13.5
14.2
11,700
0.510
0.170
1,000
22.9
13.9
14.6
15,700
0.380
0.126
500
23.5
14.5
15.2
23,000
0.260
0.086
250
24.0
15.0
15.7
34,200
0.175
0.058
100
24.8
15.8
16.5
57,000
0.105
0.035
50
25.3
16.3
17.0
80,000
0.075
0.025
20
26.0
17.0
17.7
136,000
0.044
0.015
10
26.4
17.4
18.1
176,000
0.034
0.011
5
27.0
18.0
18.7
250,000
0.024
0.008
2
27.3
18.3
19.0
333,000
0.018