| PSØ21
- Digital Amplifier for Strain Gages The
Oscillator PSØ21
offers 3 possibilities to add a reference clock: - Internal
oscillator driver for 32 kHz Oscillator
- Internal
oscillator driver for high-speed reference 2 - 5 MHz
- External
reference input
When
working in measurement range 2 (recommended) the high-speed clock is preferred.
It is used - as
predivider for the TDC measuring unit
- and
for setting the cycle time
Principally
we recommend using the clock driver of the PSØ21 with an external resonator
or quartz. The clock is used as predivider for the TDC measuring unit. Its quality
has direct influence on the quality of the measurement result. Low jitter and
short-time stability are the important properties. When using the microcontroller's
clock additional noise making the measurement quality worse is propable. The
quality needed for the resonant circuit is about 500 - 1000. From there it is
sufficient to use a ceramic resonator. A quartz oscillator is not necessary, it
produces no better results and has a longer settling time. |